LV Septum Pacing Shown Safe, Feasible Alternative for Antibradycardia Pacing

Summary

Left ventricular septum pacing is feasible and may be an alternative site for antibradycardia pacing that avoids some of the disadvantages associated with biventricular pacing. This article reviews the early investigational research and a feasibility study supporting this potential new approach.

  • antibradycardia
  • left ventricular septum pacing
  • right ventricular apex
  • LVS pacing
  • RVA pacing
  • interventricular septum
  • IVS
  • hemodynamic measurements
  • interventional techniques & devices

The use of the left ventricular (LV) septum as an alternative site for pacing was shown to be feasible, safe, and effective, without the disadvantages associated with biventricular pacing, according to Kevin Vernooy, MD, PhD, Maastricht University Medical Center, Maastricht, The Netherlands. Work conducted by his research group, which studies LV septum (LVS) pacing, was presented at the European Heart Rhythm Association/Cardiostim Inventors Awards session.

Conventional right ventricular apex (RVA) pacing is not ideal for most patients, because it leads to LV dyssynchrony and can result in LV dysfunction. In LVS pacing, the lead is placed using the same transvenous routes as the traditional right ventricular pacing lead, but a lead with an extended helix is placed through the interventricular septum (IVS) to pace the left side of the IVS.

Prof Prinzen’s research group showed in experimental work that LVS pacing was associated with less electrical dyssynchrony and more synchronous contraction [Mills RW et al. Circ Arrhythm Electrophysiol. 2009].

The feasibility study presented in this session included 10 patients with sinus node dysfunction who were not dependent on ventricular pacing. The patients had an average age of 72 years; one-half were women; the average LV ejection fraction was 59%; and IVS thickness was 9 mm. All patients were in sinus rhythm with a narrow QRS complex. After the right atrial lead was placed, an angiogram of the right ventricular was performed for close visualization. A catheter delivery system (Medtronic C315 or C304) was used to position an adapted Medtronic 3830 lead with an extended helix (Medtronic 09066) perpendicular to the IVS to provide LVS. A hemodynamic (LVdp/dtmax) pacing protocol was then performed 10 beats per minute above intrinsic sinus rate at the LVS, right ventricular septum (RVS; ring of LVS lead), and RVA.

LVS lead implantation was successful in all patients on the first attempt. During the course of the study, the time required to place the lead decreased from 90 to 12 minutes. A progressive decrease was also seen for total length of the procedure (237 to 83 minutes) and total fluoroscopy time (44 to 10 minutes). The last procedures were completed without intercardiac echocardiography.

No LVS lead-related complications occurred during the periprocedure period or at 6 months. Lead stability and other pacing parameters were not clinically different at 6 months vs baseline (Table 1).

Table 1.

Key Outcomes of Left Ventricular Septum Feasibility Study

The average QRS duration at 6 months was significantly shorter during LVS (144 milliseconds), as compared with RVA (172 milliseconds) and RVS (165 milliseconds) pacing (P < .02).

Hemodynamic measurements revealed that in healthy hearts, RVA pacing reduced LVdp/dtmax by 7.1% and by 6.9% with RVS as compared with normal intrinsic activation. With LVS pacing, LVdp/dtmax increased significantly vs RVA and RVS pacing.

In conclusion, during long-term follow-up, the LVS lead remained electrically and mechanically stable, and hemodynamic measures support LVS pacing to be preferable over RVA pacing. These researchers suggest that LVS pacing has the potential to be a better approach than RVA pacing in patients who need pacing for bradycardia.